Satelec Curing Lights Specifications | A-dec, Inc.

Satelec Curing Lights Specifications

Satelec Mini LED ScanWave curing light:

  • LED source: 4
  • Wavelength range: 390 - 510nm
  • Central wavelength:  405 / 440 / 460 and 480nm
  • Intensity: 1500 mW/cm² ± 10% (Light guide 7.5 mm); 2200 mW/cm² ± 10% (Light guide 5.5 mm)
  • Maximum exposure time: 60 seconds (pulse mode at 1500 mW/cm² ± 10%)

Handpiece

  • Weight: 185 g / 6.6oz
  • Dimensions: Ø26 x 210mm
  • Operation: Constant service
  • Classification: Type B
  • Protection: Fusible 1.5 A T Fu1(not accessible) 125 V
  • Protective index: IPX0

Laser ring

  • Wavelengt: 650 nm
  • Intensity: < 1 mW
  • Class: 2M
  • Exposure time:As long as user presses the button

 

Satelec Mini LED curing light:

  • Factory installation and streamlined integration with A-dec 500
  • Ergonomic handpiece design
  • Single LED generates 1250 mW/cm² with a 7.5 mm light guide and 2000 mW/cm² with a 5.5 mm light guide
  • Emits light in a wide spectrum: 420 nm to 480 nm
  • No ultraviolet or infrared spectrum, eliminating abnormal temperature increases and reducing the possibility of eye damage
  • Weight: 5.64 oz/160 g

Unit

  • Size: Ø 23 x 200 mm
  • Weight: 190 g
  • Classification: ordinary, type B, intermittent service
  • Working voltage: 100-240 V AC, 50-60 Hz
  • Classification: II, IP40

Base

  • Voltage: 12 V DC
  • Protection: 2 A earthed fuse
  • Classification: permanent service

Battery

  • Type: Lithium-Ion
  • Capacity: 2,000 mAh

Optical Specifications

  • Wavelength: 420-480 nm
  • Power: 1,100 mW/cm²